TB 9-6625-2331-35
SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment
Required. Table 2 identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286, AN/GSM-287, or AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI.
5. Accessories Required. The
accessories required for this
calibration are
common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure.
Table 2. Minimum Specifications of
Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
Range: 105 to 125 V ac
General Radio, Type
Accuracy: 1%
W10MT3AS3 (7910809) or
Ridge, Model 9020A (9020A)
or Ridge, Model 9020F (9020F)
CALIBRATOR
ACV: 1.416 to 6.372 Vrms @ 100 kHz
John Fluke, Model 5700A/CT
DCV: 4 to 40 V dc
(p/o MIS-35947)
Accuracy: .275%
Range: 10 MHz
Tracor, Model 527E (MIS-
METER
Resolution: 1 part in 10-10 per day
10318)
Range: -5.15 to +12.1 V dc
John Fluke, Model 8840A/AF-
Accuracy: 0.2%
05/09 (AN/GSM-64D)
Range: 1 MHz to 2 GHz
(SG-1207/U)
Amplitude: -27 to -11 dBm
Accuracy: 3%
TIME/FREQUENCY
Frequency: 10 MHz
Autek Systems Corp. Model
WORKSTATION
Accuracy: 1.25 parts in 10-10 per day
620 (MIS-38946)
TUNABLE ACTIVE FILTER
50 kHz low pass
KrohnHite, Model 3940 (3940)
SECTION III
6. Preliminary Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration process. Personnel should become familiar with the entire bulletin before
beginning the calibration.