TB 9-6625-2331-35
Table 5. Volt Max/Min - Continued
Calibrator
Test instrument
Amplitude
Input
Function
Min
Max
DC
-40.0 V
B
VMin
-40.84
-39.16
DC
40.0 V
B
VMax
39.16
40.84
DC
40.0 V
B
VMin
39.16
40.84
3
100 kHz
1.414 Vrms (4 Vpp)
B
VMax/Min
3.756
4.244
1
100 kHz
1.414 Vrms (4 Vpp)
A
VMax/Min
3.756
4.244
100 kHz
6.372 Vrms (18 Vpp)
A
VMax/Min
16.16
19.84
1
100 kHz
6.372 Vrms (18 Vpp)
B
VMax/Min
16.16
19.84
Press SWAP A↔B key.
1
Remove the tunable active filter from the connection.
2
3 Press MATH key, FUNCTION ? ? keys to select (K*X+L)/M, and ENTER.
b. Adjustments No adjustments can be made.
11. Trig
Level A&B Outputs
a. Performance Check
(1) Connect the TI TRIG LEVEL A OUT (rear panel) to the multimeter inputs.
(2) Press TI keys as listed in (a) through (d) below:
(a)
LOCAL/PRESET.
(b)
TRIGGER LEVEL SET A.
(c)
Use the DATA ENTRY keys to enter SET A levels listed in table 6 below.
(d)
ENTER.
(3) Verify that the multimeter indicates within limits listed in table 6 below.
(4) Repeat technique of (2)(b), (c), (d) and (3) above for remaining rows in table 6
Test instrument
Min
Max
SET A
SET B
(Vdc)
(Vdc)
+5.00 V
--------------------
4.720
5.280
-5.00 V
--------------------
-5.280
-4.720
0.00 V
--------------------
-0.030
+0.030
--------------------1
+5.00 V
4.720
5.280
--------------------
-5.00 V
-5.280
-4.720
--------------------
0.00 V
-0.030
+0.030
1Move
multimeter connection from TRIG LEVEL A OUT to TRIG LEVEL B OUT (rear panel)
(5) Disconnect equipment setup.
b. Adjustments No adjustments can be made.