TB 9-6625-2332-40
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
was used as the prime data source in compiling these instructions. The equipment being
calibrated will be referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. None.
b. Time and Technique. The time required for this calibration is approximately 6
hours, using the dc and low frequency and microwave technique.
2. Forms, Records, and Reports. Forms, records, and reports required for calibration
personnel at all levels are prescribed by TB 750-25.
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specifications
9210 Main Frame
Period accuracy
Range: 1 nsec to 500 msec
Accuracy: 0.5% of programmed value + 0.2 nsec
Pulse width accuracy
Range: 1 nsec to 500 msec
0.5% of programmed value + 0 nsec
Accuracy:
Pulse delay accuracy
Range: 0 nsec to 500 msec
0.5% of programmed value + 1 nsec
Accuracy:
0.035% of programmed value + 5 psec
RMS jitter
100 mV
Accuracy:
External trigger
Trigger Slope: Positive or negative
Double Pulse: When double pulse is set to on, two pulses are
produced for each trigger pulse.
9211 Module
Range: 5 V
Amplitude
Accuracy: 1% of programmed value +1% of amplitude +40 mV
V
into 50
ariable risetime/falltime
10% to 90% point
10% of programmed value + 300 psec
Accuracy:
9215 Module
Range: 15 V
Amplitude
Accuracy: 1% of programmed value + 1% of amplitude + 5 mV
V
into 50
ariable risetime/falltime
Range: 10% to 90% point
20% of programmed value + 300 psec
Accuracy: