SECTION III. CALIBRATION PROCESS
Figure 2. Termination check.

Calibration Procedure For Pulse Generator (Model 9210 Mod 200) Page Navigation
  1    2    3  4  5    6    7    8    9  
img
. . . .
TB 9-6625-2332-40
Figure 1. Test instrument configuration.
b. Connect TI to a 115 V ac power source.
c. Press POWER key to on and allow at least 15 minutes for TI to stabilize.
8. Output Amplitude
a. Performance Check
(1) On TI press RECALL SETUP pushbutton and select STANDARD key on the
crt.
(2) Press TI keys as indicated in (a) thru (c) below.
MORE.
(a)
CAL.
(b)
EXECUTE.
(c)
(3) Wait for TI to finish self calibration.
(4) Press TI keys as indicated in (a) thru (c) below.
MORE.
(a)
SELF TEST.
(b)
EXECUTE.
(c)
(5) Wait for TI to finish self test.
(6) Connect equipment as shown in figure 2 below.
(7) Setup multimeter to measure ohms.
5