Table 34. KIR Characteristics
Table 35. KIT Characteristics

Calibration Procedure For Transponder Test Set Model An/Upm-155 Page Navigation
  44    45    46    47    48  49  50    51    52    53    54  
img
. .
TB 9-6625-2337-24
Table 34. KIR Characteristics - Continued
TI
Oscilloscope
Test description
Settings/connection
Measurement type
Minimum
Maximum
GTC TRIG RT
--------------------------------------
Rise Time
---------------
100 ns
Channel 2
GTC TRIG FT
--------------------------------------
Fall Time
---------------
200 ns
Channel 2
GTC TRIG
--------------------------------------
Overshoot
---------------
5
OVERSHOOT
Channel 2
GTC TRIG 0.8 μS
--------------------------------------
Width
600 ns
1.0 μs
PW
Channel 2
GTC TRIG DLY
--------------------------------------
Delta time channel 1
371.5 μs
372.5 μs
to channel 2
Move connection from TI KIR
ISLS TRIG 4.5V
V Top
3.5 V
5.5 V
GTC TRIG OUT to TI KIR
LVL
Channel 2
ISLS TRIG OUT.
ISLS TRIG RT
--------------------------------------
Rise Time
---------------
200 ns
Channel 2
ISLS TRIG FT
--------------------------------------
Fall Time
---------------
200 ns
Channel 2
ISLS TRIG
--------------------------------------
Overshoot
---------------
5%
OVERSHOOT
Channel 2
ISLS TRIG PW
--------------------------------------
Width
100 ns
1.5 μs
Channel 2
CHVID - ISLS
Move the connection from the TI
Delta time channel 1
5.8 μs
6.0 μs
TRIGGERS M4 PRE OUT to
to channel 2
DLY
the KIR CH VID OUT.
(6) Reduce all outputs to minimum and disconnect equipment setup.
b. Adjustments. No adjustments can be made.
21. M4 KIT Simulator
a. Performance Check
(1) Connect as follows:
(a) Connect TI VAR PULSE OUTPUT to pulse generator Ext Input using a
75  feedthrough termination.
(b) Connect pulse generator (9211) OUTPUT to TI KIT EN TRIG IN using a
50  feedthrough termination.
(c) Connect TI IFF VIDEO CHAL/TAG output to KIT CH VID IN using a
75  feedthrough termination.
(d) Connect KIT REPLY OUT to oscilloscope channel 1 input using a 93
feedthrough termination.
(2) Press TI keys as listed in (a) through (g) below:
Press FUNC, ENTR keys.
(a)
Use the arrow keys to highlight GENERAL MENUS and press ENTR key.
(b)
2, ENTR (CHALLENGES).
(c)
Use the arrow keys to highlight M2:.
(d)
ENTR (OFF).
(e)
49