TB 9-6625-2337-24
(f) Use the arrow keys to highlight M4:.
(g) ENTR, (WORD A).
(3) Set the pulse generator to produce a 1.5 s wide, 1.5 V pulse with 10 s period,
trigger mode to external, and slope to negative.
(4) Oscilloscope will indicate within limits listed in first row of table 35.
(5) Repeat technique of (2) through (4) above using remaining setting and limits
TI
Test description
Settings/connection
Measurement type
Minimum
Maximum
KIT REPLY OUT
--------------------------------------
Pulse count
3
3
External
channel 1
KIT P1-P2
--------------------------------------
Delta time pulse 1 to
1.725 μs
1.775 μs
External
pulse 2 leading
edges
KIT P2-P3
--------------------------------------
Delta time pulse 2 to
1.725 μs
1.775 μs
External
pulse 3 leading
edges
KIT REPLY 4.5 V
--------------------------------------
V top
3.5 V
5.5 V
LVL
External
--------
KIT REPLY
--------------------------------------
Overshoot
5%
OVERSHOOT
External
KIT REPLY PW
--------------------------------------
Width
420 ns
480 ns
External
KIT REPLY OUT
Turn pulse generator output off
Pulse count
3
3
Internal
and remove cable from 9211
channel 1
output to TI KIT EN TRIG IN.
Disconnect cable from pulse
to TI KIT EN TRIG IN using a
75 feedthrough termination.
6
Press FUNC,
, ENTR (VAR PULSES)
Use the arrow keys to highlight
VP1:PW:
0.5, ENTR (0.5 s)
Use the arrow keys to highlight
VP1:DLY:
177.4, ENTR (177.4 s)
5 ress FUNC,
P
, ENTR (TIMING, VIDEO
LEVELS)
Use the arrow keys to highlight
CHALLENGE VIDEO:
5, ENTR (5.0 V)
KIT P1-P2
--------------------------------------
Delta time pulse 1 to
1.725 μs
1.775 μs
Internal
pulse 2 leading
edges
50