Residual FM
Figure 2. Input attenuator accuracy equipment setup.

Calibration Procedure For Spectrum Analyzers An/Usm-677 (Model E4407B-H76) And Models E4407B, E4407B-E57, E4407B-H57 And E4407B-1D51Dr Page Navigation
  9    10    11    12    13  14  15    16    17    18    19  
img
.
TB 9-6625-2339-24
NOTE
If displayed trace is not approximately 5 divisions below the
reference level, press Sweep, Sweep (Cont) keys and then
FREQUENCY key. Adjust the RPG knob to place displayed
trace approximately 5 divisions below reference level. Press
Single.
(10) Press Peak Search (or Search) key then [Pk-Pk Search] key.
(11) Record absolute value of Mkr1  amplitude indication.
(12) Multiply value recorded in (6) above by value recorded in (11) above. Result will
be ≤ 150 Hz for model E4407B and ≤ 100 Hz for all other models.
NOTE
If calibrating model E4407B stop here and proceed to
paragraph 12.  For all other models complete steps (13)
through (19) below.
(13) Press TI keys as listed in (a) through (o) below:
(a) Preset.
(b) [Factory Preset] (if it is displayed).
(c) Sweep, [Points 401].
(d) FREQUENCY, 1, [GHz].
(e) SPAN, 1, [MHz].
(f) AMPLITUDE, [Ref Level], 9, [dBm].
(g) [Scale/Div], 2, [dB].
(h) Peak Search (or Search).
(i) SPAN, [Span Zoom], 5, [kHz]. Wait for the SPAN 5 kHz message to appear.
(j)  BW/Avg, [Res BW], 1, 0, [Hz] (Man).
(k) SPAN, 1, 0, 0, [Hz].
(l)  Peak Search (or Search), Marker , [ Marker Ref Level], Marker, [Off].
(m) Single. Wait for sweep to finish.
(n) Peak Search (or Search).
(o) [Meas Tools], [Delta].
(14) Adjust TI RPG knob ccw until the Mkr1  amplitude indicates within -9.7 and -10.3 dB.
(15) Press TI keys as listed in (a) through (g) below:
(a)
Marker, [Normal].
Marker, [MkrCF].
(b)
(c)
Single.
(d)
BW/Avg, [Video BW], 1, 0, [Hz] (Man).
(e)
SPAN, [Zero Span].
(f)
Sweep, [Sweep Time], 2, 0, [ms].
(g)
Single.
14