Residual FM - Continued
Reference Level Accuracy

Calibration Procedure For Spectrum Analyzers An/Usm-677 (Model E4407B-H76) And Models E4407B, E4407B-E57, E4407B-H57 And E4407B-1D51Dr Page Navigation
  10    11    12    13    14  15  16    17    18    19    20  
img
. . . .
TB 9-6625-2339-24
NOTE
If displayed trace is not approximately 5 divisions below the
reference level, press Sweep, Sweep (Cont) key and then
FREQUENCY key. Adjust the RPG knob to place displayed trace
approximately 5 divisions below reference level. Press Single.
(16) Press Peak Search (or Search) key then [Pk-Pk Search] key.
(17) Record absolute value of Mkr1  amplitude indication.
(18) Multiply value recorded in (17) above by 0.426 Hz/dB. Result will be ≤ 2 Hz.
(19) Set all outputs to minimum and disconnect equipment setup.
b. Adjustments. None.
12. Input Attenuator Accuracy
a. Performance Check
(1) Connect equipment as shown in figure 2.
Figure 2. Input attenuator accuracy equipment setup.
(2) Set RF reference source output frequency to 50 MHz and output amplitude to
60 dBm.
(3) Press TI keys as listed in (a) through (k) below:
(a) Preset.
(b) [Factory Preset] (if it is displayed).
(c) System, [Alignments], [Auto Align], [Off].