TB 9-6625-2339-24
RF reference source
Test instrument
Mkr1 indications
Amplitude
Ref Level
settings
settings
(dB)
(dBm)
(dBm)
Min
Max
-2
-10
-9.7
-10.3
-12
-20
-19.7
-20.3
-22
-30
-29.7
-30.3
-32
-40
-39.7
-40.3
-42
-50
-49.5
-50.5
-52
-60
-59.5
-60.5
-62
-70
-69.5
-70.5
-72
-80
-79.5
-80.5
1
-82
-90
-89.3
-90.7
All models except E4407B.
1
(27) Set RF reference source amplitude to +8 dBm.
(28) Press TI keys as listed in (a) through (g) below:
(a) AMPLITUDE, [Ref Level], 0, [dBm].
(b) SPAN, 1, 5, 0, [Hz].
(c) BW/Avg, [Res BW], 1, 0, [Hz].
(d) [Video BW], 1, [Hz].
(e) Single.
(f) Peak Search (or Search).
(g) Marker, [Delta].
(29) Set RF reference source amplitude to -2 dBm.
(30) Press TI AMPLITUDE, [Ref Level], 1, 0, [dBm] keys.
(31) Press TI Single key and wait for completion of sweep.
(32) Press TI Peak Search (or Search) key. TI Mkr1 indication will be within
(33) Repeat technique of (29) through (32) above for RF reference source amplitude
and TI Ref Level settings listed in table 14.
RF reference source
Test instrument
Mkr1 indications
Amplitude
REF LEVEL
settings
settings
(dB)
(dBm)
(dBm)
Min
Max
-2
-10
-9.7
-10.3
-12
-20
-19.7
-20.3
-22
-30
-29.7
-30.3
-32
-40
-39.7
-40.3
-42
-50
-49.5
-50.5
-52
-60
-59.5
-60.5
-62
-70
-69.5
-70.5
-72
-80
-79.5
-80.5
-901
-82
-89.3
-90.7
All models except E4407B.
1
20