• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Resolution Bandwidth Accuracy
Table 18. 6 dB Resolution Bandwidth Accuracy

Calibration Procedure For Spectrum Analyzers An/Usm-677 (Model E4407B-H76) And Models E4407B, E4407B-E57, E4407B-H57 And E4407B-1D51Dr Page Navigation
  19    20    21    22    23  24  25    26    27    28    29  
img
. . . . . .
TB 9-6625-2339-24
(f) SPAN, 7, ., 5, [MHz].
(g) AMPLITUDE, [Scale/Div], 1, [dB].
(h) [More], [Y Axis Units] (or Amptd Units), [dBm].
(i) BW/Avg, [Res BW], 5, [MHz].
(j)  [Video BW], 3, 0, [Hz].
(4) Press TI Peak Search (or Search), [Meas Tools], [Mkr  CF].
(5) Adjust RF reference source amplitude to position the signal for a marker
amplitude reading of -5 dBm +/- 0.2 dB.
(6) Set RF reference source amplitude steps to 3 dB.
(7) Press TI Peak Search (or Search), Marker, [Delta].
(8) Step RF reference source amplitude up 3 dB.
(9) Press TI Marker key, then lower the marker frequency by adjusting the RPG
knob (ccw) until the marker delta amplitude is 0 +/- 0.05 dB.
(10) Record the marker frequency readout as the lower 3 dB frequency in table 17.
(11) Raise the TI marker frequency by adjusting the RPG knob (cw) until the marker
delta amplitude is 0 +/- 0.05 dB.
(12) Record the marker frequency readout as the upper 3 dB frequency in table 17.
(13) Subtract the lower 3 dB frequency recorded in (10) above from the upper 3 dB
frequency recorded in (12) above. The result will be within the limits specified in table 17.
(14) Step RF reference source amplitude down 3 dB.
(15) Press TI Marker, [Normal] keys.
(16) Repeat technique of (3) (f) and (i) and (4) through (15) above for the TI BW and SPAN
settings listed in table 17. TI Mkr1  indications will be within limits specified in table 17.
Table 17. 3 dB Resolution Bandwidth Accuracy
Test instrument
Mkr1  indications
SPAN
BW
settings
settings
(MHz)
Lower
Upper
marker
marker
Min
Max
(MHz)
(MHz)
frequency
frequency
(MHz)
(MHz)
7.5
5
3.5
6.5
4.5
3
2.55
3.45
1.5
1
0.85
1.15
0.450
0.300
0.255
0.345
0.150
0.100
0.085
0.115
0.045
0.030
0.0255
0.0345
0.015
0.010
0.0085
0.0115
0.0045
0.003
0.00255
0.00345
0.0015
0.001
0.00085
0.00115
(17) Set RF reference source amplitude to -7 dBm.
(18) Press TI BW/Avg, [Res BW], 1, 2, 0, [kHz] and SPAN, 1, 8, 0, [kHz] keys.
(19) Press TI Peak Search (or Search), [Meas Tools], [Mkr  CF].
24