Figure 4. Frequency response equipment setup.
Table 20. Frequency Response < 50MHz

Calibration Procedure For Spectrum Analyzers An/Usm-677 (Model E4407B-H76) And Models E4407B, E4407B-E57, E4407B-H57 And E4407B-1D51Dr Page Navigation
  22    23    24    25    26  27  28    29    30    31    32  
img
. . .
TB 9-6625-2339-24
(8) Press TI Peak Search (or Search) key.
(9) Record TI Mkr1 indication.
(10) Set synthesized signal generator to 18 GHz.
(11) Press TI FREQUENCY, 1, 8, [GHz] keys.
(12) Adjust synthesized signal generator power level for a measuring receiver
indication of -10 +/- 0.05 dBm.
(13) Press TI Single key.
(14) Press TI Peak Search (or Search) key.
(15) TI Mkr1 indication will be within limits listed for appropriate model in table 19 of
the indication recorded in (9) above.
(16) Repeat (10) through (15) above for the remaining synthesized signal generator
frequency and TI FREQUENCY settings listed in table 19.
Table 19. Frequency Response > 50 MHz
Synthesized signal
Test instrument
generator
Model E4407B
All other models
Frequency settings
Mkr1 limits
Mkr1 limits
FREQUENCY
(dB)
(dB)
18.0 GHz
18.0 GHz
2.0
2.0
17.5 GHz
17.5 GHz
2.0
2.0
17.0 GHz
17.0 GHz
2.0
2.0
16.5 GHz
16.5 GHz
2.0
2.0
16.0 GHz
16.0 GHz
2.0
2.0
15.5 GHz
15.5 GHz
2.0
2.0
15.0 GHz
15.0 GHz
2.0
2.0
14.5 GHz
14.5 GHz
2.0
2.0
14.0 GHz
14.0 GHz
2.0
2.0
13.5 GHz
13.5 GHz
2.0
2.0
13.0 GHz
13.0 GHz
2.0
2.0
12.5 GHz
12.5 GHz
2.0
2.0
12.0 GHz
12.0 GHz
2.0
2.0
11.5 GHz
11.5 GHz
2.0
2.0
11.0 GHz
11.0 GHz
2.0
2.0
10.5 GHz
10.5 GHz
2.0
2.0
10.0 GHz
10.0 GHz
2.0
2.0
9.5 GHz
9.5 GHz
2.0
2.0
9.0 GHz
9.0 GHz
2.0
2.0
8.5 GHz
8.5 GHz
2.0
2.0
8.0 GHz
8.0 GHz
2.0
2.0
7.5 GHz
7.5 GHz
2.0
2.0
7.0 GHz
7.0 GHz
2.0
2.0
6.5 GHz
6.5 GHz
1.5
2.0
6.0 GHz
6.0 GHz
1.5
2.0
5.5 GHz
5.5 GHz
1.5
2.0
5.0 GHz
5.0 GHz
1.5
2.0
4.5 GHz
4.5 GHz
1.5
2.0
27


Privacy Statement