Table 20. Frequency Response < 50MHz
Displayed Average Noise - Continued

Calibration Procedure For Spectrum Analyzers An/Usm-677 (Model E4407B-H76) And Models E4407B, E4407B-E57, E4407B-H57 And E4407B-1D51Dr Page Navigation
  24    25    26    27    28  29  30    31    32    33    34  
TB 9-6625-2339-24
(29) Set all outputs to minimum and disconnect equipment setup.
b. Adjustments. None.
18. Displayed Average Noise
a. Performance Check
(1) Connect TI AMPTD REF OUT to INPUT 50.
(2) Press TI keys as listed in (a) through (m) below:
(a) Preset.
(b) [Factory Preset] (if it is displayed).
(c) Sweep, [Points 401].
(d) Input/Output (or Input), [Amptd Ref Out] (On).
(e) FREQUENCY, 5, 0, [MHz].
(f) SPAN, 2, [kHz].
(g) AMPLITUDE, [Ref Level], 2, 0, [dBm].
(h) [Attenuation], 1, 0, [dB].
(i) BW/Avg, [Res BW], 1, [kHz].
(j)  [Video BW], 1, [kHz].
(k) Det/Demod, [Detector], [Sample], ◄Return.
(l) Single.
(m) Peak Search (or Search).
(3) Record the TI MARKER amplitude reading as reading 1.
(4) Press TI keys as listed in (a) through (f) below:
(a) AMPLITUDE, [Attenuation], 0, [dB].
(b) SPAN, 2, 0, [kHz].
(c) BW/Avg, [Res BW], 1, [kHz].
(d) [Video BW], 3, 0, [Hz].
(e) Single.
(f) Peak Search (or Search).
(5) Record the TI MARKER amplitude reading as reading 2.
(6) Calculate reference level offset by subtracting reading 2 from reading 1. If the
calculated reference level offset is greater than 0.05 dB or less than 0.05 dB, record as Ref
Level Offset (1 kHz RBW) value, otherwise, enter 0.
(7) Press TI keys as listed in (a) through (f) below:
(a) AMPLITUDE, [Attenuation], 0, [dB].
(b) SPAN, 5, 0, 0, [Hz].
(c) BW/Avg, [Res BW], 1, 0, [Hz].
(d) [Video BW], 1, [Hz].
(e) Single.
(f) Peak Search (or Search).
(8) Record the TI MARKER amplitude reading as reading 3.