TB 9-6625-2339-24
Figure 5. Probe power connector.
Table 24. Probe Power Connector Voltages
Test instrument connector
Multimeter indication (Vdc)
voltage
Min
Max
12.6 V
13.86
11.3
+15  V
13.95
16.05
(10) Disconnect equipment setup.
(11) Press TI power STANDBY pushbutton.
(12) Disconnect TI power cord.
(13) Reinstall TI inner shield and outer case.
(14) Connect TI power cord.
(15) Press TI power ON pushbutton.
b. Adjustments. None.
21. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
34