Figure 2. Input attenuator accuracy equipment setup.
Figure 3. Reference level accuracy equipment setup.

Calibration Procedure For Spectrum Analyzers An/Usm-677 (Model E4407B-H76) And Models E4407B, E4407B-E57, E4407B-H57 And E4407B-1D51Dr Page Navigation
  11    12    13    14    15  16  17    18    19    20    21  
img
. . . . . .
TB 9-6625-2339-24
(d)
Sweep, [Points 401].
(e)
FREQUENCY, 5, 0, [MHz].
(f)
SPAN, 1, 0, 0, [kHz].
AMPLITUDE, [Ref Level], 5, 5, [dBm].
(g)
(h)
[Attenuation], 1, 0, [dB].
(i)
[Scale/Div], 2, [dB].
(j)
BW/Avg, [Res BW], 3, 0, [kHz].
(k)
[Video BW], 1, 0, 0, [Hz].
(4) Verify signal peak 2 to 4 divisions below TI reference level.
(5) Press TI Peak Search (or Search), Marker, [Delta] keys.
(6) Set RF reference source to 70 dBm.
(7) Press TI AMPLITUDE, [Ref Level], 6, 5, [dBm] then [Attenuation], 0, [dB].
(8) Press TI Single, then Peak Search (or Search). TI Mkr1  indication will be
within limits specified in first row of table 10.
(9) Repeat (6) through (8) above for remaining RF reference source amplitude, TI
Ref Level and Atten settings listed in table 10.
Table 10. Input Attenuator Accuracy
RF reference
source
Test instrument
Mkr1  indications
Amplitude
Ref Level
Atten
settings
settings
settings
(dB)
(dBm)
(dBm)
(dB)
Min
Max
10.3
9.7
-70
-65
0
5.3
4.7
-65
-60
5
-55
-50
15
4.7
5.3
-50
-45
20
9.7
10.3
-45
-40
25
14.65
15.35
-40
-35
30
19.6
20.4
-35
-30
35
24.55
25.45
-30
-25
40
29.5
30.5
-25
-20
45
34.45
35.55
-20
-15
50
39.4
40.6
-15
-10
55
44.35
45.65
-10
-5
60
49.3
50.7
-5
0
65
54.25
55.75
(10) Set all outputs to minimum and disconnect equipment setup.
b. Adjustments. None.
13. Reference Level Accuracy
a. Performance Check
(1) Connect equipment as shown in figure 3.