(5) Press TI keys as listed in (a) through (c) below:
(a) Single (wait for completion of sweep).
(b) Peak Search.
(c) Marker, [Delta].
(6) Set synthesized signal generator frequency to 2.700 GHz.
(7) Press TI Single key, wait for completion of sweep then press Peak Search key.
(8) TI Mkr1 indication will be within limits specified in table 8.
(9) Repeat technique of (3) (d) through (8) above for TI Start, Stop and synthesized
signal generator frequency settings listed in table 8.
Test instrument
Mkr1
Frequency
Frequency
settings in (4)
settings in (6)
FREQUENCY
FREQUENCY
indications
stop frequency
start frequency
above
above
Min
Max
300
MHz
2.7
GHz
0
Hz
3
GHz
2.370 GHz
2.430 GHz
20
MHz 100
MHz
10
MHz
110
MHz
79.0 MHz
81.0 MHz
10.01
MHz
10.09
MHz
10
MHz
10.1
MHz
79.0
kHz
81.0
kHz
810
MHz 890
MHz
800
MHz
900
MHz
79.0 MHz
81.0 MHz
800.01
MHz 800.09
MHz
800
MHz
800.1
MHz
79.0
kHz
81.0
kHz
1.410 GHz
1.490 GHz
1.4
GHz
1.5
GHz
79.0 MHz
81.0 MHz
1.49901 GHz
1.49909 GHz
1.499 GHz
1.4991 GHz
79.0
kHz
81.0
kHz
(10) Set all outputs to minimum and disconnect equipment setup.
b. Adjustments. None.
a. Performance Check
(1) Connect TI 10 MHz REF OUT to RF reference source REF FREQUENCY INPUT
connector (rear panel).
(2) Connect RF reference source leveling head output connector to TI INPUT 50.
(3) Set RF reference source for an output frequency of 1 GHz at an output level of 0 dBm.
(4) Press TI keys as listed in (a) through (x) below:
(a) Preset.
(b) [Factory Preset] (if it is displayed).
(d) Sweep, [Points 401].
(e) SPAN, 1, 0, [MHz].
(g) [More], [Y Axis Units] (or Amptd Units), [dBm].
(h) Peak Search (or Search).
(j) SPAN, 5, 0, [kHz].
(k) BW/Avg, [Res BW], 1, [kHz] (Man).