Bandwidth Accuracy and Selectivity
Bandwidth Switching

Calibration Procedure For Spectrum Analyzer, Model 8566B Page Navigation
  16    17    18    19    20  21  22    23    24    25    26  
img
. .
TB 9-6625-2340-24
SCALE LIN.
(f)
SHIFT.
(g)
AUTOA.
(h)
REFERENCE LEVEL.
(i)
(27) Adjust DATA knob to position trace at (or just below) reference level (top)
graticule line.
(28) Press TI keys as listed in (a) through (d) below:
MARKER NORMAL.
(a)
MARKER PEAK SEARCH.
(b)
MARKER CF.
(c)
MARKER .
(d)
(29) Adjust TI DATA knob to place marker down the positive going side of the
displayed signal to the 3 dB point.
(30) Adjust A4A9R65 (fig. 8) for a MKR
indication of 5 kHz while maintaining
marker at the 3 dB point using TI DATA knob (R).
key.
(31) Press TI MARKER
(32) Adjust TI DATA knob to set marker to the 3 dB point on the negative going
side of the displayed signal. There are now two markers; one on each side of the signal at
the 3 dB points.
(33) TI MKR  indication should be 10 1.0 kHz.
(34) Press TI keys as listed in (a) through (i) below:
MARKER OFF.
(a)
CENTER FREQUENCY, 100, MHz.
(b)
FREQUENCY SPAN, 5, kHz.
(c)
RES BW, 3, kHz.
(d)
REFERENCE LEVEL, 10, dBm.
(e)
SCALE LIN.
(f)
SHIFT.
(g)
AUTOA.
(h)
REFERENCE LEVEL.
(i)
(35) Adjust DATA knob to position trace at (or just below) reference level (top)
graticule line.
(36) Press TI keys as listed in (a) through (d) below:
MARKER NORMAL.
(a)
MARKER PEAK SEARCH.
(b)
MARKER CF.
(c)
MARKER .
(d)
(37) Adjust TI DATA knob to place marker down the positive going side of the
displayed signal to the 3 dB point.
(38) Adjust A4A9R66 (fig. 8) for a MKR  indication of 1.5 kHz while maintaining
marker at the 3 dB point using TI DATA knob (R).
key.
(39) Press TI MARKER
21