Bandwidth Switching
Figure 11. A4A7 board - test points.

Calibration Procedure For Spectrum Analyzer, Model 8566B Page Navigation
  18    19    20    21    22  23  24    25    26    27    28  
img
. . . .
TB 9-6625-2340-24
(3) Set TI LINE switch to ON.
(4) Press TI keys as listed in (a) through (h) below:
2-22 GHz.
(a)
CENTER FREQUENCY, 100, MHz.
(b)
FREQUENCY SPAN, 0, Hz.
(c)
ATTEN, 0, dB.
(d)
RES BW, 1, kHz.
(e)
SCALE LIN.
(f)
MARKER .
(g)
RES BW, 1, MHz.
(h)
(5) Adjust A4A6R29 (fig. 9) to align markers on TI display. MKR
level should
indicate 1.00X.
Figure 9. A4A6 board.
(6) Press TI keys as listed in (a) and (b) below:
(a) 2-22 GHz.
(b) RECALL, 9.
(7) Adjust TI FREQ ZERO for maximum trace amplitude.
NOTE
The pictured filter (fig. 10) is required to continue with this
adjustment procedure.
Figure 10. Filter.
(8) Press TI keys as listed in (a) through (e) below: