Bandwidth Switching - Continued
Figure 17. A4A4 board.

Calibration Procedure For Spectrum Analyzer, Model 8566B Page Navigation
  23    24    25    26    27  28  29    30    31    32    33  
img
. . . . . .
TB 9-6625-2340-24
Figure 15. A4A5 board.
(59) Multimeter should indicate between 8.5 and 11.0 V dc. If measurement is out of
tolerance, adjust A4A5R2 for a multimeter indication of 9.5 0.1 V dc (R).
(60) Connect equipment as shown in figure 16 below.
(61) Set both attenuators to 0 dB.
(62) Disconnect 97 (white/violet) cable from A4A8J1 (fig. 13).
(63) Disconnect 89 (gray/white) cable from A4A6J1 (fig. 9).
(64) Connect 97 (white/violet) cable to A4A6J1 (fig. 9).
Figure 16. Attenuator connection.
28