Bandwidth Accuracy and Selectivity - Continued
Figure 9. A4A6 board.

Calibration Procedure For Spectrum Analyzer, Model 8566B Page Navigation
  17    18    19    20    21  22  23    24    25    26    27  
img
. . . . . .
TB 9-6625-2340-24
(40) Adjust TI DATA knob to set marker to the 3 dB point on the negative going
side of the displayed signal. There are now two markers; one on each side of the signal at
the 3 dB points.
indication should be 3.0 0.30 kHz.
(41) TI MKR
(42) Press TI 2-22 GHz key.
(43) Replace TI top cover.
13. Bandwidth Switching
a. Performance Check
(1) Press TI 2-22 GHz key.
(2) Connect TI CAL OUTPUT to TI RF INPUT.
(3) Press TI keys as listed in (a) through (f) below:
CENTER FREQUENCY, 100, MHz.
(a)
FREQUENCY SPAN, 5, MHz.
(b)
REFERENCE LEVEL, 8, dBm.
(c)
RES BW, 1, MHz.
(d)
LOG ENTER dB / DIV, 1, dB.
(e)
MARKER PEAK SEARCH, .
(f)
(4) Press TI MARKER PEAK SEARCH key.
(5) Read the amplitude deviation from the marker
indication.  If indication
displayed is not within limits specified in first row of table 9, perform b below.
(6) Repeat technique of (3) (b), (d), (4) and (5) above for remaining TI RES BW and
FREQUENCY SPAN key settings in table 9. If indications displayed are not within limits
specified in table 9, perform b below.
Table 9. Bandwidth Switching Accuracy
Test instrument
Deviation
RES BW
FREQUENCY
(dB)
SPAN
key
key
Min
Max
1  MHz
5  MHz
0 (ref)
0 (ref)
1.00
3  MHz
5  MHz
1.00
0.50
300  kHz
5  MHz
0.50
0.50
100  kHz
500  kHz
0.50
0.50
30  kHz
500  kHz
0.50
0.50
10  kHz
50  kHz
0.50
0.50
3  kHz
50  kHz
0.50
0.50
1  kHz
10  kHz
0.50
0.50
300  Hz
1  kHz
0.50
0.50
100  Hz
1  kHz
0.50
0.80
30
Hz
200  Hz
0.80
2.00
10
Hz
100  Hz
2.00
b. Adjustments
(1) Set TI LINE switch to STANDBY.
(2) Remove TI top cover.
22