TB 9-6625-2340-35
(5) Press TI keys as listed in (a) and (b) below:
(a) CENTER FREQUENCY, 40, MHz.
(b) FREQUENCY SPAN, 20, kHz.
frequency).
(7) Press TI MARKER NORMAL, PEAK SEARCH.
(9) Press TI PEAK SEARCH.
(10) TI MARKER ∆ frequency indication on TI display will indicate within limits
(11) Repeat technique of steps (5)(b) through (10) above for remaining signal
generator No. 1 frequencies and TI frequency spans listed in table 5. TI MARKER ∆
frequency indications will indicate within limits specified in table 5.
Signal generator
Test instrument
Indications
Indications
Low frequency
High frequency
Frequency
(∆)
(∆)
(MHz)
(MHz)
span
Min
Max
39.992
40.008
20 kHz
15.84 kHz
16.16 kHz
39.980
40.020
50 kHz
39.60 kHz
40.40 kHz
39.940
40.060
150 kHz
118.80 kHz
121.20 kHz
39.920
40.080
200 kHz
158.4
kHz
161.6
kHz
39.600
40.400
1 MHz
792.00 kHz
808.00 kHz
39.200
40.800
2 MHz
1.584 MHz
1.616 MHz
37.600
42.400
6 MHz
4.656 MHz
4.944 MHz
36.000
44.000
10 MHz
7.76 MHz
8.240 MHz
20.000
60.000
50 MHz
38.80 Hz
41.20 MHz
(12) Disconnect TI RF INPUT from signal generator No. 1 RF OUTPUT.
(13) Connect TI RF INPUT to signal generator No. 3 RF OUTPUT.
(14) Disconnect TI 10 MHz OUT from signal generator No. 1 EXT REF INPUT.
(15) Connect TI 10 MHz OUT (rear panel) to signal generator No. 3 FREQ
STANDARD EXT (rear panel).
(16) Set the signal generator No. 3 for an output frequency of 4 GHz at a -10 dBm
level.
(17) Press TI keys as listed in (a) and (b) below:
(a) CENTER FREQUENCY, 4, GHz.
(b) FREQUENCY SPAN, 500, MHz.