TB 9-6625-2341-24
Table 1. Calibration Description - Continued
Test instrument
Performance specifications1
parameters
Range: 30, 300, 3 k, 30 k, 300 k, 3 M, and 30 M ohm
Accuracy: 30 ohm, (2.5% +25 counts), 300 ohm to 30 M ohm, (0.5% + 5
counts)
Frequency
Range: 1 Hz to 5 MHz
Accuracy: (0.5% +2 counts)
1
All specifications are +/-(% of indication +counts).
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286; AN/GSM-287; or AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI. Where the four-to-one ratio
cannot be met, the actual accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure. The following peculiar accessories are also required for this
calibration: 50 : feedthrough, BNC plug to BNC jack, Tektronix, Type 011-0049-01(011-
0049-01).
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
CALIBRATOR
Range: 100 mV ac 200 V ac
Fluke, Model 5720A (5720A) (p/o
Accuracy:
MIS-35947); w amplifier, Fluke
5 Hz - 20 Hz
+/-1.25%
5725A/AR (5725A/AR)
20 Hz - 50 Hz
+/-0.875%
50 Hz - 60 Hz
+/-1.25%
60 Hz 20 kHz
+/-0.875%
20 kHz 1 MHz
+/-1.25%
1 MHz 5 MHz
+/-3.125%
Volt range:
Accuracy:
.1 - 300 V dc
+/-0.166%
Ohm range:
Accuracy:
10 ohm
+/-
1.25%
100 - 10 M ohm
+/-0.25%
Accuracy:
1 MHz
+/-0.175%