TB 9-6625-2341-24
WARNING
HIGH VOLTAGE is used or exposed during the performance of
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
a. Remove TI from PM 9083 holster.
b. Connect PM 8907/803 AC/DC ADAPTOR supplied with TI to a 115 V ac power source.
c. Insert PM 8907/803 AC/DC ADAPTOR low voltage plug into TI right side POWER
ADAPTOR connector.
d. Press TI ON/OFF key to ON.
NOTE
The TI uses the same Input Circuitry for the SCOPE and
METER modes. The remaining CH A deflection coefficients will
be checked when the METER mode is verified.
a. Performance Check
(1) Reset by pressing pushbutton sequence (a) through (m) below:
(a) Display SCOPE/METER/:/EXT.mV and etc. Press (F1) SCOPE to
display/highlight SCOPE.
NOTE
If the SCOPE mode is already active, press the SUB MENU
key. Press (F1) CLOSE. If "The red INPUT A and grey
INPUT B BNC jacks are being used as inputs now", then press
(F1) CLOSE.
(b) USER OPTIONS.
(c) (F1) MORE OPTION.
(d) Toggle
or
to display/highlight RESET MENU.
(e) (F5) SELECT ITEM.
(f) Toggle
or
to display/highlight RESET SCOPEMETER.
(g) (F5) SELECT ITEM.
(h) (If all ScopeMeter settings have been set to original setting then (F1)
CLOSE), which will display/highlight RESET SCOPEMETER.
(i) Toggle
or
to display/highlight RESET SCOPE.
(j) (F5) SELECT ITEM.