TB 9-6625-2341-24
Table 4. Scope Vertical Gain - Continued
Test instrument
Output volt
Voltage error
settings at 1 kHz
(%)
Range settings
(V p-p)
Min
Max
(V/Div)
50 m
.2
-3
3
100 m
.4
-3
3
200 m
.8
-3
3
500 m
2
-3
3
1
4
-3
3
2
8
-3
3
5
20
-3
3
10
40
-3
3
20
80
-3
3
1
100
-3
3
50
2
100
-3
3
100
1
Adjust 5820A edit amplitude output controls for 2 vertical divisions.
2
Adjust 5820A edit amplitude output controls for 1 vertical divisions.
(11) (F1) CLOSE.
b. Adjustments. Complete adjustments are in paragraph 16 (R).
a. Performance Check
(1) Reset by pressing pushbutton sequence (a) through (l) below:
(a) USER OPTIONS.
(b) (F1) MORE OPTION.
(c) Toggle
or
to display/highlight RESET MENU.
(d) (F5) SELECT ITEM.
(e) Toggle
or
to display/highlight RESET SCOPEMETER.
(f) (F5) SELECT ITEM.
(g) (If all ScopeMeter settings have been set to original setting then (F1)
CLOSE), which will display/highlight RESET SCOPEMETER.
(h) Toggle
or
to display/highlight RESET SCOPE.
(i) (F5) SELECT ITEM.
(j) (For original settings, ScopeMeter needs to be selected. Change to SCOPE
now? (F1) YES then (F1) CLOSE) or (if all ScopeMeter settings have been set to original
setting then (F1) CLOSE), which will display/highlight RESET SCOPE.
(k) (F1) CLOSE which will display/highlight RESET MENU.
(l) (F1) CLOSE which will display/highlight CONTR.
(2) Connect equipment as shown in figure 3 below: