Scope Bandwidth - Continued
Scope Trigger - Continued

Calibration Procedure For Scopemeter, Model 105B Page Navigation
  8    9    10    11    12  13  14    15    16    17    18  
TB 9-6625-2341-24
10. Scope Trigger
a. Performance Check
(1) Reset by pressing pushbuttons sequence (a) through (l) below:
(a) USER OPTIONS.
(b) (F1) MORE OPTION.
(c) Toggle
or
to display/highlight RESET MENU.
(d) (F5) SELECT ITEM.
(e) Toggle
or
to display/highlight RESET SCOPEMETER.
(f) (F5) SELECT ITEM.
(g) (If all ScopeMeter settings have been set to original setting then (F1)
CLOSE), which will display/highlight RESET SCOPEMETER.
(h) Toggle
or
to display/highlight RESET SCOPE.
(i) (F5) SELECT ITEM.
(j) (For original settings, SCOPE mode needs to be selected.  Change to
SCOPE mode now? (F1) YES then (F1) CLOSE) or (if all ScopeMeter settings have been
set to original setting then (F1) CLOSE), which will display/highlight RESET SCOPE.
(k) (F1) CLOSE which will display/highlight RESET MENU.
(l) (F1) CLOSE which will display/highlight CONTR.
(2) RESET oscilloscope calibrator.
(3) Press pushbutton sequence as listed in (a) through (e) below:
(a) "Input A" to off by pressing (i) through (vii) below:
(i)  INPUT A.
(ii) (F1) MORE INPUT A.
(iii) Toggle
or
to display/highlight INPUT A.
(iv) (F5) SELECT ITEM.
(v) Toggle
or
to display/highlight off.
(vi) (F5) SELECT ITEM.
(vii) (F1) CLOSE.
(b) "Input B" probe to 1:1 by pressing (i) through (ix) below:
(i)
INPUT B to display/highlight DISPLAY INPUT B.
(ii) (F1) MORE INPUT B.
(iii) Toggle
or
to display/highlight PROBE B MENU.
(iv) (F5) SELECT ITEM which will display/highlight PROBE on INPUT B.
(v) (F5) SELECT ITEM which will display/highlight (1:1, 10:1, 100:1,
1000:1, etc.).