TB 9-6625-2341-24
(16)Set oscilloscope calibrator LEVSINE output to produce a 0.13 V p-p at 100 MHz.
(17) Using oscilloscope calibrator, adjust edit dial amplitude to 2.1 vertical divisions
on TI. If TI signal is not triggered, perform b below.
(18) Set range by toggling (mV RANGE V) on left side, under INPUT A pushbutton
to display/highlight A 200 mV DC 1:1.
(19) Set oscilloscope calibrator LEVSINE output to produce a .3 V p-p at 10 MHz.
(20) Using oscilloscope calibrator, adjust edit dial amplitude to 1.4 vertical divisions
on TI. If TI signal is not triggered, perform b below.
(21) Set oscilloscope calibrator to standby.
b. Adjustments. Complete adjustments are in paragraph 16 (R).
Scope Time base
a. Performance Check
(1) Reset by pressing pushbuttons sequence (a) through (l) below:
(a) USER OPTIONS.
(b) (F1) MORE OPTION.
(c) Toggle
or
to display/highlight RESET MENU.
(d) (F5) SELECT ITEM.
(e) Toggle
or
to display/highlight RESET SCOPEMETER.
(f) (F5) SELECT ITEM.
(g) (If all ScopeMeter settings have been set to original setting then (F1)
CLOSE), which will display/highlight RESET SCOPEMETER.
to display/highlight RESET SCOPE.
(h) Toggle
or
(i) (F5) SELECT ITEM.
(j) (For original settings, SCOPE mode needs to be selected. Change to
SCOPE mode now? (F1) YES then (F1) CLOSE) or (if all ScopeMeter settings have been
set to original setting then (F1) CLOSE), which will display/highlight RESET SCOPE.
(k) (F1) CLOSE which will display/highlight RESET MENU.
(l) (F1) CLOSE which will display/highlight CONTR.
(2) RESET oscilloscope calibrator.
(3) Press pushbutton sequence as listed in (a) through (d) below:
(a) "Input B" to off by pressing (i) through (vii) below:
(i)
INPUT B.
(ii) (F1) MORE INPUT B.
(iii) Toggle
or
to display/highlight INPUT B.
(iv) (F5) SELECT ITEM.
(v) Toggle
or
to display/highlight off.