Meter Dc Voltage
Table 5. Meter DC Volts

Calibration Procedure For Scopemeter, Model 105B Page Navigation
  13    14    15    16    17  18  19    20    21    22    23  
img
.
TB 9-6625-2341-24
CALIBRATOR
CH A
COM
OUTPUT HI & LO
TI
Figure 5. Meter dc voltage.
(2) Reset by pressing pushbuttons sequence (a) through (k) below:
(a) SCOPEMETER/MAIN MENU to display SCOPE/METER/:/EXT.mV and etc.
(b) (F2) METER to display/highlight METER.
(c) USER OPTIONS should display MORE OPTIONS.
(d) (F1) MORE OPTION.
(e) Toggle
or
to display/highlight RESET MENU.
(f) (F5) SELECT ITEM.
(g) Toggle
or
to display/highlight RESET METER.
(h) (F5) SELECT ITEM.
(i) All meter settings have been set to original setting then (F1) CLOSE which
will display/highlight RESET METER.
(j) (F1) CLOSE will display/highlight RESET MENU.
(k) (F1) CLOSE.
(3) Set "Input A" probe to 1:1 by pressing (a) through (g) below:
(a) INPUT A.
(b) (F5) PROBE A MENU.
(c) Toggle
or
to display/highlight PROBE on INPUT A.
(d) (F5) SELECT ITEM which will display/highlight (1:1, 10:1, 100:1,
1000:1, etc.).
(e) Toggle
or
to display/highlight 1:1.