Figure 6. Meter ac voltage.
Table 8. EXT.mV DC

Calibration Procedure For Scopemeter, Model 105B Page Navigation
  18    19    20    21    22  23  24    25    26    27    28  
img
. . .
TB 9-6625-2341-24
14. Meter EXT.mV
a. Performance Check
(1) Connect equipment as shown in figure 7 below:
(2) Select reset by pressing pushbuttons sequence (a) through (c) below:
(a) SCOPEMETER/MAIN MENU to display SCOPE/METER/:/EXT.mV
and etc.
(b) (F5) EXT.mV to display (The banana jacks are being used as inputs now).
(c) (F1) CLOSE to display/highlight EXT.mV.
CALIBRATOR
Figure 7. Meter EXT.mV.
(3) Select reset by pressing pushbuttons sequence (a) through (i) below:
(a) USER OPTIONS.
(b) (F1) MORE OPTION.
(c) Toggle
or
to display/highlight RESET MENU.
(d) (F5) SELECT ITEM.
(e) Toggle
or
to display/highlight RESET EXTERNAL mV.
(f) (F5) SELECT ITEM.
(g) (If all EXT.mV settings have been set to original setting then press (F1)
CLOSE), which will display/highlight RESET EXTERNAL mV or (for original settings,
24