• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Table 8. EXT.mV DC
Meter Resistance - Continued

Calibration Procedure For Scopemeter, Model 105B Page Navigation
  20    21    22    23    24  25  26    27    28    29    30  
img
. . . .
TB 9-6625-2341-24
(d) (F5) SELECT ITEM.
(e) Toggle
or
to display/highlight Vrms AC.
(f) (F5) SELECT ITEM.
(g) (F1) CLOSE).
(h) SCOPEMETER/MAIN MENU to display SCOPE/METER/:/EXT.mV and etc.
(8) Perform steps as listed in (a) through (d) below for each row in table 9.
(a)  Under INPUT A pushbutton on left side, toggle (mV RANGE V) to display
TI Range settings.
(b)
Set calibrator to Output volt settings at 60 Hz.
(c) If TI voltage indications are not within specified limits, perform b below.
(d) Set calibrator to standby.
Table 9. EXT.mV AC
Calibrator
Test instrument
Voltage indication
(V ac)
Output volt settings
Range settings
Min
Max
(V ac)
(V ac)
.250
300 m
243.5 m
256.5 m
2
3
1.945
2.055
(9) Reset calibrator.
b. Adjustments. Complete adjustments are in paragraph 16 (R).
15. Meter Resistance
a. Performance Check
(1) Select resistance by pressing pushbutton (F3) : to display/highlight :.
(2) Select reset by pressing pushbuttons sequence (a) through (i) below:
(a) USER OPTIONS.
(b) (F1) MORE OPTION.
(c) Toggle
or
to display/highlight RESET MENU.
(d) (F5) SELECT ITEM.
(e) Toggle
or
to display/highlight RESET OHM.
(f) (F5) SELECT ITEM.
(g) (If all OHM settings have been set to original setting then press (F1)
CLOSE), which will display/highlight RESET OHM or (for original settings, OHM needs
to be selected. Change to OHM mode now press (F1) YES then (F1) CLOSE).
(h) (F1) CLOSE will display/highlight RESET MENU.
(i) (F1) CLOSE.
26


Privacy Statement
Press Release
Contact

© Copyright Integrated Publishing, Inc.. All Rights Reserved.