TB 9-6625-2341-24
(3) Press SCOPEMETER/MAIN to display/highlight SCOPE/METER/:/EXT.mV
and etc.
(4) Perform steps as listed in (a) through (c) below for each row in table 10.
(a) Under INPUT A pushbutton on left side, toggle (mV RANGE V) to display
TI Range settings.
(b) Set calibrator to Output ohm settings.
(c) If TI Ohms indications are not within specified limits, perform b below.
Calibrator
Test instrument
Output ohm settings
Range settings
Ohms indication
(:)
(:)
(:)
0
30
<0.25
0
300
<0.5
(5) Perform steps as listed in (a) through (d) below for each row in table 11.
(a) Under INPUT A pushbutton on left side, toggle (mV RANGE V) to display
TI Range settings.
(b) Set calibrator to Output ohm settings.
(c) Adjust calibrator percent of error dial until calibrator reads exact indication on TI.
(d) If calibrator is not within Error indications, perform b below.
Test instrument
Calibrator
Error indications (%)
Range settings
Output ohm settings
(:)
(:)
Min
Measured
Min
1
30
10
-5
+5
300
100
-1
+1
3k
1k
-1
+1
30 k
10 k
-1
+1
100 k2
300 k
-1
+1
3M
1M
-1
+1
30 M
10 M
-1
+1
1
Set calibrator 2 Wire COMP to on.
2
Set calibrator 2 Wire COMP to off.
(6) Reset calibrator.
(7) Disconnect all equipment.
b. Adjustments. Complete adjustments are in paragraph 16 (R).