Figure 9. Input short.
Calibration Constants Restoration - Continued

Calibration Procedure For Scopemeter, Model 105B Page Navigation
  25    26    27    28    29  30  31    32    33    34    35  
img
.
TB 9-6625-2341-24
OSCILLOSCOPE CALIBRATOR
Figure 10. Attenuation and gain.
(14) Select pulse response of *1 attenuation by pressing pushbuttons sequence (a)
through (d) below:
(a) Toggle
to display/highlight (
CAL 6).
or
or
(b) Set oscilloscope calibrator to produce a square wave, 1 kHz at 300 mV P-P.
(c) Press (F5) READY, then wait till display/highlight READY is
not highlighted.
(d) Oscilloscope calibrator to STBY.
(15) Select pulse response of *10 attenuation by pressing pushbuttons sequence (a)
through (d) below:
(a) Toggle
to display/highlight (
CAL 7).
or
or
(b) Set oscilloscope calibrator to produce a square wave, 1 kHz at 3 V p-p.
(c) Press (F5) READY, then wait till display/highlight READY is
not highlighted.
(d) Oscilloscope calibrator to STBY.
(16) Select pulse response of *100 attenuation by pressing pushbuttons sequence
(a) through (d) below:
(a) Toggle
to display/highlight (
CAL 8).
or
or
(b) Set oscilloscope calibrator to produce a square wave, 1 kHz at 20 V p-p.
(c ) P r e s s ( F 5 ) R E A D Y , t h e n w a i t t i l l d i s p l a y / h i g h l i g h t R E A D Y i s
n o t highlighted.
(d) Oscilloscope calibrator to STBY.
31