TB 9-6625-2341-24
(30) Set channel A and B, 50 percent and 90 percent trigger level by pressing
pushbuttons sequence (a) through (i) below:
(a) Toggle
to display/highlight (
CAL 20).
or
or
(b) Set calibrator to produce an output of 10 kHz at 707 mV p-p.
(c)
Press (F5) READY, then wait till display/highlight READY is not highlighted.
(d) Toggle
to display/highlight (
CAL 21).
or
or
(e)
Press (F5) READY, then wait till display/highlight READY is not highlighted.
(f)
Press
to display/highlight (
CAL 22).
or
or
(g) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(h) Press
to display/highlight (
CAL 23).
or
or
(i)
Press (F5) READY, then wait till display/highlight READY is not highlighted.
(31)
Set calibrator to standby.
(32)
Connect equipment as shown in figure 12 below:
CALIBRATOR
(33)
Select external trigger level by pressing pushbuttons sequence (a) through (c) below:
(a) Toggle
to display/highlight (
CAL 24).
or
or
(b) Set calibrator to operate.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
34