Calibration Constants Restoration - Continued
Figure 14. 10:1 probe for INPUT A.

Calibration Procedure For Scopemeter, Model 105B Page Navigation
  32    33    34    35    36  37  38    39  
img
. .
TB 9-6625-2341-24
(d) Displayed SCOPE/METER/:/EXT.mV and etc.
(e)
Press SPECIAL FUNCT.
(f)
(F2) GENERATE MENU.
(g) Toggle
to display/highlight (F2) GENERATOR OUTPUT: ON/OFF.
or
(h) (F5) SELECT ITEM to display/highlight GENERATOR OUTPUT: ON.
(i)
(F1) CLOSE.
(j)
SCOPEMETER/MAIN MENU to display SCOPE/METER/:/EXT.mV and etc.
(k) Press pushbutton BLACKLIGHT (sun shaped pushbutton just below
SPECIAL FUNCT key) with AUTO SET together, then release.
(l)
(F4) CALIBRATE SCOPEMETER.
(m) Displayed are the numbers of calibration memories free. Press (F1) CLOSE.
(n) (F3) METER.
(o)
Set calibrator to standby.
(p) Remove + 12 volts from calibrator to TI.
(59)
Connect equipment as shown in figure 13 below:
CALIBRATOR
Figure 13. Current ramp.
NOTE
CAL numbering sequence might not be in step.
(60) Select current ramp by pressing pushbuttons sequence (a) through (c) below:
(a) Toggle
to display/highlight (
CAL 18).
or
or
(b) Set calibrator two wire resistance to produce 100 ohms.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
Reset calibrator.
(61)
38