TB 9-6625-2341-24
(24) Select gain for 2 V by pressing pushbuttons sequence (a) through (d) below:
(a) Toggle
to display/highlight (
CAL 16).
or
or
(b) Set oscilloscope calibrator to produce a square wave, 1 kHz at 10 V p-p.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(25)
Select gain for 20 V by pressing pushbuttons sequence (a) through (d) below:
(a) Toggle
to display/highlight (
CAL 17).
or
or
(b) Set oscilloscope calibrator to produce a square wave, 1 kHz at 100 V p-p.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(26)
Shift gain *1 mode by pressing pushbuttons sequence (a) through (d) below:
(a) Toggle
to display/highlight (
CAL 18).
or
or
(b) Set oscilloscope calibrator to produce a square wave, 1 kHz at 200 mV p-p.
(c)
Press (F5) READY, then wait till display/highlight READY is not highlighted.
(27)
Shift gain /8 mode by pressing pushbuttons sequence (a) through (c) below:
(a) Toggle
to display/highlight (
CAL 19).
or
or
(b) Set oscilloscope calibrator to produce a square wave, 1 kHz at 20 mV p-p.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(28)
Reset oscilloscope calibrator.
(29)
Connect equipment as shown in figure 11 below:
CALIBRATOR
33