Figure 10. Attenuation and gain.
Figure 11. Trigger level.

Calibration Procedure For Scopemeter, Model 105B Page Navigation
  26    27    28    29    30  31  32    33    34    35    36  
TB 9-6625-2341-24
(17)  Select pulse response of *1000 attenuation by pressing pushbuttons sequence (a)
through (d) below:
(a) Toggle
to display/highlight (
CAL 9).
or
or
(b) Set oscilloscope calibrator to produce a square wave, 1 kHz at 50 V p-p.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(d) Oscilloscope calibrator to STBY.
(18)
Select gain for 5 mV by pressing pushbuttons sequence (a) through (d) below:
(a) Toggle
to display/highlight (
CAL 10).
or
or
(b) Set oscilloscope calibrator to produce a square wave, 1 kHz at 20 mV p-p.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(d) Oscilloscope calibrator to STBY.
(19)
Select gain for 10mV by pressing pushbuttons sequence (a) through (d) below:
(a) Toggle
to display/highlight (
CAL 11).
or
or
(b) Set oscilloscope calibrator to produce a square wave, 1 kHz at 50 mV p-p.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(d) Oscilloscope calibrator to STBY.
(20)
Select gain for 20mV by pressing pushbuttons sequence (a) through (d) below:
(a) Toggle
to display/highlight (
CAL 12).
or
or
(b) Set oscilloscope calibrator to produce a square wave, 1 kHz at 100 mV p-p.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(d) Oscilloscope calibrator to STBY.
(21)
Select gain for 50mV by pressing pushbuttons sequence (a) through (d) below:
(a) Toggle
to display/highlight (
CAL 13).
or
or
(b) Set oscilloscope calibrator to produce a square wave, 1 kHz at 200 mV p-p.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(d) Oscilloscope calibrator to STBY.
(22)
Select gain for 100mV by pressing pushbuttons sequence (a) through (d) below:
(a) Toggle
to display/highlight (
CAL 14).
or
or
(b) Set oscilloscope calibrator to produce a square wave, 1 kHz at 500 mV p-p.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(d) Oscilloscope calibrator to STBY.
(23)
Select gain for 200mV by pressing pushbuttons sequence (a) through (d) below:
(a) Toggle
to display/highlight (
CAL 15).
or
or
(b) Set oscilloscope calibrator to produce a square wave, 1 kHz at 1 V p-p.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.