Calibration Constants Restoration - Continued
Calibration Constants Restoration - Continued

Calibration Procedure For Scopemeter, Model 105B Page Navigation
  30    31    32    33    34  35  36    37    38    39  
img
.
TB 9-6625-2341-24
(48)  Select Channel A, 300 mV, 1 V, 3 V, 30 V, 100 V, 300 V, range/gain by pressing
pushbuttons sequence (a) through (x) below:
(a) Toggle
to display/highlight (
CAL 4).
or
or
(b) Set calibrator to produce a 300 mV dc signal.
(c)
Press (F5) READY, then wait till display/highlight READY is not highlighted.
(d) Set calibrator to STBY.
(e)
Toggle
to display/highlight (
CAL 23).
or
or
(f)
Set calibrator to produce a 1 V dc signal.
(g) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(h) Set calibrator to STBY.
(i)
Toggle
to display/highlight (
CAL 5).
or
or
(j)
Set calibrator to produce a 3 V dc signal.
(k) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(l)
Set calibrator to STBY.
(m) Toggle
to display/highlight (
CAL 6).
or
or
(n) Set calibrator to produce a 30 V dc signal.
(o)
Press (F5) READY, then wait till display/highlight READY is not highlighted.
(p) Set calibrator to STBY.
(q) Toggle
to display/highlight (
CAL 24).
or
or
(r)
Set calibrator to produce a 100 V dc signal.
(s)
Press (F5) READY, then wait till display/highlight READY is not highlighted.
(t)
Set calibrator to STBY.
(u) Toggle
to display/highlight (
CAL 7).
or
or
(v) Set calibrator to produce a 300 V dc signal.
(w) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(x) Reset calibrator.
(49) Connect equipment as shown in figure 7.
(50) Select EXT mV, 300 mV, 3 V, range/gain by pressing pushbuttons sequence (a)
through (g) below:
(a) Toggle
to display/highlight (
CAL 8).
or
or
(b) Set calibrator to produce a 300 mV dc signal.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(d) Set calibrator to STBY.
(e) Toggle
to display/highlight (
CAL 9).
or
or
(f)
Set calibrator to produce a 3 V dc signal.
(g) Press (F5) READY, then wait till display/highlight READY is not highlighted.
36