TB 9-6625-2341-24
a. Performance Check
(1) Reset by pressing pushbuttons sequence (a) through (i) below:
(a) USER OPTIONS.
(b) (F1) MORE OPTION.
(c) Toggle
or
to display/highlight RESET MENU.
(d) (F5) SELECT ITEM.
(e) Toggle
or
to display/highlight RESET METER.
(f) (F5) SELECT ITEM.
(g) All meter settings have been set to original setting then (F1) CLOSE which
will display/highlight RESET METER.
(h) (F1) CLOSE will display/highlight RESET MENU.
(i) (F1) CLOSE.
(2) RESET calibrator.
(3) Set measure to VAC by pressing (a) through (d) below:
(a) MEASURE MENU to display/highlight (V, Hz, dB, TIME).
(b) (F2) V.
(c) Toggle
or
to display/highlight Vrms AC METER.
(d) (F5) SELECT ITEM.
(4) Set "Input A" probe to 1:1 by pressing (a) through (g) below:
(a) INPUT A.
(b) (F5) PROBE A MENU.
(c) Toggle
or
to display/highlight PROBE on INPUT A.
(d) (F5) SELECT ITEM which will display/highlight (1:1, 10:1, 100:1,
1000:1, etc.).
(e)
or
to display/highlight 1:1.
(f) (F5) SELECT ITEM.
(g) (F1) CLOSE which will display/highlight CONTR.
(5) Select display change by pressing (a) through (h) below:
(a) SUB MENU.
(b) (F1) MORE METER.
(c) Toggle
or
to display/highlight READINGS on DISPLAY.
(d) (F5) SELECT ITEM which will display/highlight (1, 2, 3, 4).
(e) Toggle
or
to display/highlight 1.
(f) (F5) SELECT ITEM.