Figure 3. Scope bandwidth input A.
Scope Bandwidth - Continued

Calibration Procedure For Scopemeter, Model 105B Page Navigation
  6    7    8    9    10  11  12    13    14    15    16  
img
. .
TB 9-6625-2341-24
(5) Set oscilloscope calibrator LEVSINE output to produce a 120 mV at 50 kHz and
adjusting edit amplitude output dial for 6 vertical divisions on TI.
(6) Set oscilloscope calibrator frequency output to produce 80 MHz.
(7) Using oscilloscope calibrator, increase frequency edit dial until frequency drops
the amplitude down to 4.2 divisions on TI. If frequency is not greater than or equal to 100
MHz, perform b below.
(8) Set oscilloscope calibrator to standby.
(9) Connect equipment as shown in figure 4 below:
50:
feedthrough
termination
OSCILLOSCOPE
CALIBRATOR
COM
CH B
SOURCE/MEASURE
EXT TRIG CHAN 1
TI
Figure 4. Scope bandwidth input B.
(10) Press pushbutton sequence as listed in (a) through (e) below:
(a) "Input A" to off by pressing (i) through (vi) below:
(i)
(F1) MORE INPUT A.
(ii) Toggle
or
to display/highlight INPUT A.
(iii) (F5) SELECT ITEM.
(iv) Toggle
or
to display/highlight off.
(v) (F5) SELECT ITEM.
(vi) (F1) CLOSE.
(b) "Input B" to on by pressing (i) through (vii) below:
(i)
INPUT B.
(ii) (F1) MORE INPUT B.
(iii) Toggle
or
to display/highlight INPUT B.
(iv) (F5) SELECT ITEM.
(v) Toggle
or
to display/highlight ON.
12