TB 9-6625-2344-35
Test instrument
Indications
Test point
(V)
(V)
Min
Max
+5.1
5.0
5.2
-5.2
-5.3
-5.1
+12.2
11.9
12.5
-12.2
-12.5
-11.9
+3.3
2.8
3.5
(8)
Disconnect equipment setup and reinstall TI cover.
b. Adjustments. None
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
ALIGNMENT PROCESS
NOTE
Self-calibration should be performed if TI fails any parameter.
key to enable graphical interface.
a. Using TI mouse, click
b. Click UTILITIES menu on top of TI display.
d. Click TI CAL MEMORY PROTECT box to uncheck it.
NOTE
Clear CAL MEMORY PROTECT to perform self calibration.
You cannot perform self calibration if this box is checked.
e. Click START key and follow instructions displayed in text box.