TB 9-6625-2344-35
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this check. DEATH ON CONTACT may result if personnel fail
to observe safety precautions.
NOTE
Do not perform power supply check if all other parameters are
within tolerance.
a. Performance Check
(1) Press TI
button to power down TI.
(2) Disconnect power cord and remove TI cover.
(3) Reconnect power cord and press TI
button to power up TI.
(5) Connect multimeter positive lead to TI +5.1 V test point (fig.1).
(6) Multimeter indication will be within limits specified in first row of table 15.
(7) Repeat technique of (4) and (5) above for remaining test points listed in table 14
below. Multimeter indications will be within limits specified.