TB 9-6625-2344-35
(43) In MEASUREMENT DEFINITIONS window, place cursor over TO EDGE #
box and left click mouse to activate percent.
(44) Using either TI keyboard or mouse, enter 101 and click OK key.
(45) In MEASUREMENT DEFINITIONS window, click CLOSE key.
(46) TI ∆ TIME MEAN indication will be within limits specified in table 11.
Test instrument
∆ time mean indication
(s)
Min
Max
3.94960
3.95040
(47) Click MEASURE menu on top of TI display.
(48) From displayed drop-down menu, click MEASUREMENT DEFINITIONS.
(49) In MEASUREMENT DEFINITIONS window, place cursor over TO EDGE #
box and left click mouse to activate keypad.
(50) Using either TI keyboard or mouse, enter 201 and click OK key.
(51) In MEASUREMENT DEFINITIONS window, click CLOSE key.
(52) TI ∆ TIME MEAN indication will be within limits specified in table 12.
Test instrument
∆ time mean indication
(s)
Min
Max
7.8992
7.9008
(53) Set function generator output to minimum and disconnect equipment setup.
b. Adjustments. None.
a. Performance Check
(1) Connect oscilloscope calibrator SOURCE/MEASURE CHAN 1-4 outputs to TI
channels 1-4 input connectors respectively.
(2) Press TI DEFAULT SETUP button.
(3) Click SETUP menu on top of TI display.
(4) From displayed drop-down menu, click ACQUISITION.
(5) In ACQUISITION SETUP window, click SAMPLING MODE EQUIVALENT
TIME and AVERAGING ENABLED box.