TB 9-6625-2344-35
(20) Click MEASURE menu on top of TI display.
(21) From displayed drop-down menu, select TIME.
(22) From TIME submenu, click DELTA TIME.
(23) In ENTER MEASUREMENT INFO window, select CHANNEL 1 in both
SOURCE 1 and SOURCE 2 boxes.
(24) In ENTER MEASUREMENT INFO window, click OK key.
(25) Click MEASURE menu on top of TI display.
(26) From displayed drop-down menu, click MEASUREMENT DEFINITIONS.
(27) In MEASUREMENT DEFINITIONS window, click DELTA TIME tab.
(28) In MEASUREMENT DEFINITIONS window, place cursor over FROM EDGE
# box and left click mouse to activate percent.
(29) Using either TI keyboard or mouse, enter +1 and click OK key.
(30) In MEASUREMENT DEFINITIONS window, click ∇ key of FROM EDGE
DIRECTION box and select RISING.
(31) In MEASUREMENT DEFINITIONS window, click ∇ key of FROM EDGE
THRESHOLD box and select MIDDLE LEVEL.
(32) In MEASUREMENT DEFINITIONS window, place cursor over TO EDGE #
box and left click mouse to activate percent.
(33) Using either TI keyboard or mouse, enter +2.
(34) In MEASUREMENT DEFINITIONS window, click ∇ key of TO EDGE
DIRECTION box and select RISING.
(35) In MEASUREMENT DEFINITIONS window, click ∇ key of TO EDGE
THRESHOLD box and select MIDDLE LEVEL.
(36) In MEASUREMENT DEFINITIONS window, click CLOSE key.
(37) TI ∆ TIME minimum and maximum indications will be within limits specified in
Test instrument
∆ time indications
(ns)
Min
Max
(≥)
(≤)
24.926
25.074
(38) Change function generator output period to 50 ns.
(39) Click SETUP menu on top of TI display.
10