TB 9-6625-2344-35
(62) From TIME submenu, click DELTA TIME.
(63) In ENTER MEASUREMENT INFO window, select CHANNEL 1 in both
SOURCE 1 and SOURCE 2 boxes.
(64) In ENTER MEASUREMENT INFO window, click OK key.
(65) TI ∆ TIME minimum and maximum indications will be within limits specified in
Test instrument
∆ time indications
Min
Max
(≥)
(≤)
997.4 ns
1.0026 s
(66) Click MEASURE menu on top of TI display.
(67) From displayed drop-down menu, click MEASUREMENT DEFINITIONS.
(68) In MEASUREMENT DEFINITIONS window, place cursor over TO EDGE #
box and left click mouse to activate percent.
(69) Using either TI keyboard or mouse, enter 6 and click OK key.
(70) In MEASUREMENT DEFINITIONS window, click CLOSE key.
(71) Click TI CLEAR ALL key in lower left corner of screen.
(72) Click MEASURE menu on top of TI display.
(73) From displayed drop-down menu, select TIME.
(74) From TIME submenu, click DELTA TIME.
(75) In ENTER MEASUREMENT INFO window, select CHANNEL 1 in both
SOURCE 1 and SOURCE 2 boxes.
(76) In ENTER MEASUREMENT INFO window, click OK key.
(77) TI ∆ TIME minimum and maximum indications will be within limits specified in
Test instrument indications
(s)
Min
Max
(≥)
(≤)
4.9972
5.0028
(78) Set function generator output to minimum.
b. Adjustments. None.