TB 9-6625-2344-35
(40) From displayed drop-down menu, click HORIZONTAL.
(41) In HORIZONTAL SETUP window, place cursor over SCALE box and left click
mouse to activate percent.
(42) Using either TI keyboard or mouse, enter 100 n.
(43) In HORIZONTAL SETUP window, place cursor over POSITION box and left
click mouse to activate percent.
(44) Using either TI keyboard or mouse, enter -11.0 n.
(45) In HORIZONTAL SETUP window, click CLOSE key.
(46) Click TI CLEAR ALL key in lower left corner of screen.
(47) Click MEASURE menu on top of TI display.
(48) From displayed drop-down menu, select TIME.
(49) From TIME submenu, click DELTA TIME.
(50) In ENTER MEASUREMENT INFO window, select CHANNEL 1 in both
SOURCE 1 and SOURCE 2 boxes.
(51) In ENTER MEASUREMENT INFO window, click OK key.
(52) TI ∆ TIME minimum and maximum indications will be within limits specified in
Test instrument
∆ time indications
(ns)
Min
Max
(≥)
(≤)
49.69
50.31
(53) Change function generator output period to 1 s.
(54) Click SETUP menu on top of TI display.
(55) From displayed drop-down menu, click HORIZONTAL.
(56) In HORIZONTAL SETUP window, place cursor over SCALE box and left click
mouse to activate percent.
(57) Using either TI keyboard or mouse, enter 1 .
(58) In HORIZONTAL SETUP window, click CLOSE key.
(59) Click TI CLEAR ALL key in lower left corner of screen.
(60) Click MEASURE menu on top of TI display.
(61) From displayed drop-down menu, select TIME.
11