TB 9-6625-2347-35
Table 1. Calibration Description - Continued
Test instrument parameters
Performance specifications
Horizontal
Range: 0.5 s/div to 0.05 s/div
A sweep timing
Accuracy: 3%
Range: (X10 mag): 50 ms/div to 5 ns/div
Accuracy: 5%
Sweep linearity
Accuracy: 3% (measured over any 2 of the center
8 divisions)
Deflection (X-Axis)
Range: 2 mV/div to 10 V/div
Accuracy: 5%
≤ 2 MHz
2 MHz to 60 MHz
A trigger sensitivity
Internal
0.4 div
1.5 div
External
50 mV
250 mV
Probe adjust
Range: 0.5 V p-p
Accuracy: 20%
EQUIPMENT REQUIREMENTS
4. Equipment
Required. Table 2 identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286, AN/GSM-287, or AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI. Where the four-to-one ratio
cannot be met, the four-to-one accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The
accessories required for this
calibration are common
usage accessories issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure. The following peculiar accessory is also required for this calibration:
standardizer, 5-80 pF.
Table 2. Minimum Specifications of
Equipment Required
Common name
Minimum use specifications
Manufacturer and model (part number)
Range: -8.64 to < 0.1 V dc
John Fluke, Model 8840A/AF-05/09
Accuracy: 0.25%
(AN/GSM-64D)
Volts out:
John Fluke, Model 5820A, MIS-38938
Range: 10 mV to 50 V
(5820A-5C-GHZ)
Accuracy: 0.75%
Time markers:
Range: 5 ns/D to 0.5 s/D
Accuracy: 0.75%
Range: 50 kHz to >100 MHz