TB 9-6625-2347-35
NOTE
To perform step below; press oscilloscope calibrator EDIT
FIELD pushbutton as required to place underline under one of
(18) Rotate oscilloscope calibrator knob below EDIT FIELD pushbutton to sweep
oscilloscope calibrator from 50 kHz to frequency limit specified in first row of table 5 while
observing displayed waveform amplitude on TI crt. Displayed waveform amplitude will be
within limits specified in first row of table 5 throughout entire frequency range sweep.
(19) Repeat technique of (15) (b) and (16) through (18) above for remaining TI
settings and oscilloscope calibrator outputs in table 5. Displayed waveform amplitude will
be within limits specified in table 5 throughout entire frequency range sweep.
Oscilloscope calibrator
Test instrument
LEVEL SINE mode output
Amplitude
Frequency sweep
VOLTS/DIV
SEC/DIV
Amplitude limits
(divisions)
20 s
≥ 4.2
12 mV
50 kHz to 50 MHz
2m
1
20 s
≥ 4.2
60 mV
50 kHz to 60 MHz 1
10 m
20 s
≥ 4.2
3.0 V
50 kHz to 60 MHz 1
.5 V
Press Set to 50 kHz blue soft pushbutton to quickly return to 50 kHz.
1
(20) Set TI VERTICAL MODE CH 1 BOTH CH 2 switch to CH 2.
(21) Ensure TI CH 2 VOLTS/DIV switch is set to 2m.
(22) Set oscilloscope calibrator for a CHAN 2, LEVEL SINE mode output of 12 mV
at a frequency of 50 kHz.
(23) Rotate oscilloscope calibrator knob below EDIT FIELD pushbutton to adjust
amplitude for 6 divisions of deflection on TI.
NOTE
To perform step below; press oscilloscope calibrator EDIT
FIELD pushbutton as required to place underline under one of
(24) Rotate oscilloscope calibrator knob below EDIT FIELD pushbutton to sweep
oscilloscope calibrator from 50 kHz to frequency limit specified in first row of table 6 while
observing displayed waveform amplitude on TI crt. Displayed waveform amplitude will be
within limits specified in first row of table 6 throughout entire frequency range sweep.
(25) Repeat technique of (21) through (24) above for remaining TI settings and
oscilloscope calibrator outputs in table 6. Displayed waveform amplitude will be within
limits specified in table 6 throughout entire frequency range sweep.
7