TB 9-6625-2365-35
Power meter
Test instrument indication limits
no. 2
Indication
Width
Period
(dBm)
Min
Max
26.5 GHz
100
ns
1.1
s
-10.0
0.080 s
0.120
s
26.5 GHz
1
s
2
s
-10.0
0.980 s
1.02
s
26.5 GHz
10
s
11
s
-10.0
9.08
s
10.02
s
26.5 GHz
100
s
101
s
-10.0
99.08
s
100.02
s
26.5 GHz
1 ms
1.001 ms
-10.0
999.08
s
1000.02
s
26.5 GHz
10 ms
10.001 ms
-10.0
9999.08
s
10000.02
s
26.5 GHz
100 ms
100.001 ms
-10.0
99999.08
s
100000.02
s
40
GHz
100
ns
1
s
-10.0
0.080 s
0.120
s
40
GHz
1
s
2
s
-10.0
0.980 s
1.02
s
40
GHz
10
s
11
s
-10.0
9.08
s
10.02
s
40
GHz
100
s
101
s
-10.0
99.08
s
100.02
s
40
GHz
1 ms
1.001 ms
-10.0
999.08
s
1000.02
s
40
GHz
10 ms
10.001 ms
-10.0
9999.08
s
10000.02
s
40
GHz
100 ms
100.001 ms
-10.0
99999.08
s
100000.02
s
b. Adjustments. None.
a. Deenergize and disconnect all equipment.