TB 9-6625-2365-35
(3) Adjust synthesized signal generator RF output controls and variable attenuator
no. 2 controls for a power meter no. 2 indication of -10 dBm.
(4) Press TI keys as listed in (a) through (e) below.
(a)
Menu Freq.
(b)
> (until Set Freq is selected).
(c)
Enter.
(d)
Menu, 2, 6, ., 5, GHz.
(e)
Return to Meas.
(5) Set synthesized signal generator for pulse modulation with a pulse width of 100 ns
and period of 500 ns.
(6) Press TI Meas Mode Burst CW key (Burst LED lit).
indication will be within limits specified in first row of table 12.
(7) Repeat technique of (2) through (6) above for frequencies and power meter no. 2
indications listed in table 12. TI burst frequency indication will be within limits specified
Table 12. Burst Frequency (
26.5 GHz to 40
GHz)
Synthesized signal
Power meter
generator
no. 2
Test instrument indication limits
Indication
(dBm)
Min
Max
26.5 GHz
-10.0
26486
26514
30 GHz
-10.0
29984
30016
34 GHz
-10.0
33982
34018
40 GHz
-10.0
39979
40021
(8) Set synthesized signal generator RF OUTPUT to off.
(9) Press TI keys as listed in (a) through (e) below.
(a)
Menu Freq.
(b)
> (until Set Freq is selected).
(c)
Enter.
(d)
Menu, 2, 6, ., 5, GHz.
(e)
Return to Meas.
(10) Set synthesized signal generator for pulse modulation with a pulse width of 100 ns
and period of 1.1 s and turn synthesized signal generator RF OUTPUT on. TI burst
width indication will be within limits specified in first row of table 13.
(11) Set synthesized signal generator RF OUTPUT to off.
(12) Repeat (9) through (11) above for synthesized signal generator frequencies, pulse
widths, periods and power meter no. 2 indications listed in table 13. TI burst width
indication will be within limits specified in table 13.
17