TB 9-6625-357-24
b. Adjustments
(1) Adjust C211 (fig. 3) until true rms voltmeter indicates at least 0.03 V (R).
(2) Adjust P203 (fig. 1) until true rms voltmeter indicates at least 0.3 V (R).
(3) Repeat a (6) through (10) above.
(4) Adjust C316 (fig. 3) until true rms voltmeter indicates 0.0631 V (R).
(5) Repeat a (6) through (13) above.
increase C316 (fig. 3), until true rms voltmeter indicates 0.708 V (R).
(8) Repeat a (6) through (16) above.
11. Maximum Attenuation and Rate of Attenuation
a. Performance Check
(1) Set selector switch to LOW PASS X100.
(2) Set CUTOFF FREQ Hz dial to 20.
b. Adjustments. No adjustments can be made.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.