Table 3. 80 to 50,000 MMF Range

Calibration Procedure For Capacitance Analyzer Zm-3/U And Zm-3/Au Page Navigation
  1    2    3  4  5    6    7    8    9  
img
.
TB 9-6625-381-24
Figure 1. Left-side - adjustment location.
CAUTION
Do not allow METER SWITCH to go to 60 VOLTS position
during the following checks; damage to TI meter will result.
(6) Adjust OPERATING VOLTAGE control for a 500 V indication on TI meter.
(7) Operate and hold METER SWITCH to 50 MA.
Multimeter will indicate
between 482 and 518 V dc. Release METER SWITCH.
(8) Adjust OPERATING VOLTAGE control for a 600 V indication on TI meter.
(9) Operate and hold METER SWITCH to 50 MA.
Multimeter will indicate
between 582 and 618 V dc. Release METER SWITCH.
(10) Adjust OPERATING VOLTAGE control for minimum indication on TI meter.
b. Adjustments. No adjustments can be made.