SECTION I. IDENTIFICATION AND DESCRIPTION

Calibration Procedure For Ohmmeters I-48, Zm-21/U; Aul, Model 1602, 679, 7676, 7676-1, 7679, 7679-1, And 210359 Page Navigation
1  2    3    4    5    6  
img
. . . . . . . . . . . . . . . . . . . . . . . . . . .
*TB 9-6625-975-24
DEPARTMENT OF THE ARMY TECHNICAL BULLETIN
CALIBRATION PROCEDURE FOR
OHMMETERS I-48( ), ZM-21( )/U;
AUL, MODEL 1602; BIDDLE, MODELS 679,
7676, 7676-1, 7679, 7679-1, AND 210359
Headquarters, Department of the Army, Washington, DC
12 December 2007
Distribution Statement A: Approved for public release; distribution is unlimited.
REPORTING OF ERRORS AND RECOMMENDING IMPROVEMENTS
You can improve this manual. If you find any mistakes or if you know of a way to improve these
procedures, please let us know.  Mail your letter or DA Form 2028 (Recommended Changes to
Publications and Blank Forms) directly to: Commander, U.S. Army Aviation and Missile Command,
ATTN: AMSAM-MMC-MA-NP, Redstone Arsenal, AL 35898-5000. A reply will be furnished to
you.
You  may also send in your comments electronically to our E-mail address:
2028@redstone.army.mil or by fax 256-842-6546/DSN 788-6546.  For the World Wide Web use:
https://amcom2028.redstone.army.mil. Instructions for sending an electronic 2028 can be found at
the back of this manual.
Paragraph
Page
SECTION
I.
IDENTIFICATION AND DESCRIPTION
Test instrument identification ............................
1
2
Forms, records, and reports.................................
2
2
Calibration description ........................................
3
2
II.
EQUIPMENT REQUIREMENTS
Equipment required .............................................
4
3
Accessories required.............................................
5
3
III.
CALIBRATION PROCESS FOR I-48( ),
ZM-21( )/U, BIDDLE, MODELS 679, 7676,
7676-1, 7679, 7679-1, AND 210359
Preliminary instructions......................................
6
4
Equipment setup ..................................................
7
4
Meter range ..........................................................
8
4
Voltage check ........................................................
9
5
Resistance measurement .....................................
10
5
Final procedure ....................................................
11
7
__________
*This bulletin supersedes TB 9-6625-975-35, dated 20 April 1988.